Design of Electrocardiograph Signal Simulator
Abstract
Medical equipment functional test and calibration is a routine activity that must be carried on periodically. Electrocardiograph (ECG) requires an ECG phantom to calibrate the function. This calibrator is commonly called ECG signal simulator. The purpose of this study is to design a simple ECG signal simulator with ten leads of signals that can be used to test ECG recorders with standard recording procedures. With the ECG signal simulator that was designed and made, the development of signal patterns can be made as needed. The normal human cardiac signal displayed on the ECGSIM software. The potential value that displayed on ECGSIM software can be extracted manually and assembled as a flash program of microcontroller, so this microcontroller will generate some digital code by each parallel port. This digital code then converted as an analog signal by DAC. The electrocardiograph signal simulator output is an analog signal that identical with each lead according to the recording method of bipolar, unipolar and precordial of ECG. This analog signal was tested using a standard ECG recorder. It is proved that the simulator is able to generate an electrical signal in accordance with the characteristics of the human cardiac signal displayed on ECGSIM software. The results of human electrocardiograph signal simulator design are a device that generates electrical signals with output specifications that correspond to the bioelectric signals of the human heart. The statistical test showed that the p-value is more than 0.05. It is mean that there is no significant difference between the design and standard. The signals pattern has met the specification of ECGSIM signal
Downloads
Copyright (c) 2020 Catur Suharinto, Anwar Budianto, Nugroho Tri Sanyoto
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Authors who publish with this journal agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution-ShareAlikel 4.0 International (CC BY-SA 4.0) that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).